RegWardenFDA facility compliance records
Recall

Class II — Hitachi America, Ltd., Particle Therapy Division

Recalling firmHitachi High-Tech America, Inc. Radiation Therapy Division
Registered facilityHITACHI HIGH-TECH AMERICA, INC. RADIATION THERAPY DIVISION (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonDuring positioning within the system's 3D to 3D matching mode while transferring CBCT images, when the Float Image Selection button is pressed, it may display CBCT images that are insufficient in number.
ProductBrand Name: PROBEAT Product Name: PROBEAT-V Model/Catalog Number: N/A Software Version: N/A Product Description: The PROBEAT-V is a proton beam irradiation system, which provides a therapeutic proton beam for clinical treatment. It is designed to deliver a proton beam with the prescribed dose, dose distribution and directed to the prescribed patient treatment site. Component: N/A
Quantity1
DistributionU.S. Nationwide distribution in the states of TN, TX, and WA.
Initiated2024-05-27
Classified2024-09-12
Recall numberZ-3127-2024
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Hitachi America, Ltd., Particle Therapy Division (Z-3127-2024).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/94928/