Class II — Palodex Group Oy

Recalling firmPaloDEx Group Oy
Registered facilityPALODEX GROUP OY (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonIt was noticed in production on 28.10.2021, that the final testing database gives a pass result for Pan ADC repeatability test even when the test result is not within the set acceptance criteria (repeatability within 4%).
ProductOrthopantomograph OP300 dental computed tomography x-ray system. The X-ray device produces conventional 2D X-ray images and Xray projection images for the reconstruction of a 3D view. The device is operated and used by qualified healthcare professionals.
Quantity103
DistributionUS Nationwide-Worldwide
Initiated2021-10-28
Classified2024-08-12
Recall numberZ-2413-2024
InitiationFDA Mandated