Class I — B Braun Medical Inc
| Recalling firm | B. BRAUN MEDICAL SAS |
|---|---|
| Registered facility | B. BRAUN MEDICAL (the firm's only registered facility) |
| Classification | Class I |
| Status | Ongoing |
| Reason | Extension of previous recall RES 92978. Pumps have the potential for upstream occlusion sensor to cease proper function due to buildup of electrostatic charge during operation triggering false downstream and upstream occlusion alarms. Additional devices have been affected. |
| Product | Infusomat Space Infusion System/ Large Volume Pump, Product Code 8713050U |
| Quantity | 14 |
| Distribution | US distribution to AL, CA, GA, IL, MI, and TX. |
| Initiated | 2024-08-07 |
| Classified | 2024-09-06 |
| Recall number | Z-2972-2024 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class I recall — B Braun Medical Inc (Z-2972-2024).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/95093/
Alerts for B. BRAUN MEDICAL SAS
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.