Class II — Westmed, Inc.
| Recalling firm | WESTMED, INC. |
|---|---|
| Registered facility | WESTMED, INC. (the firm's only registered facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | Due to connector within the circuit flex extender not meeting specifications resulting in a low connection force and as a result, the connector may leak or disconnect and may cause life threating conditions or death. |
| Product | 1) Flex Extenders, 5.2" x 15mm OD x 15mm ID REF S6411P 2) Flex Extender, 5.2" x 15mm OD x 22/15mm ID REF 7 5132 The Westmed Flex Extenders are non invasive, single use/disposable, large bore, corrugated plastic tubes that provide a flexible, low resistance lightweight connection between breathing circuits and patient interface (tracheal tube, tracheostomy tube, or mask). |
| Quantity | 50,050 units |
| Distribution | U.S.: AK, CA, CO, DC, FL, GA, IL, IN, KY, MA, MD, ME, MI, MN, MO, NC, NE, NH, NJ, NV, NY, OH, OR, PA, SC, TX, VA, VT, WA, and WV O.U.S.: N/A |
| Initiated | 2024-09-27 |
| Classified | 2024-10-30 |
| Recall number | Z-0263-2025 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Westmed, Inc. (Z-0263-2025).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/95453/
Alerts for WESTMED, INC.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?