RegWardenFDA facility compliance records
Recall

Class II — ImaCor Inc.

Recalling firmImaCor Inc
Registered facilityIMACOR, Inc (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonEpoxy used to seal the edge of the transducer into its paddle exceeded specification. This variability could attenuate the ultrasound signal with a potential impact on the quality of images obtained from the probes.
ProductClariTEE Probe- Miniature transesophageal echo ultrasound probe for use with ImaCor ultrasound diagnostic platforms. Model/Catalog Number: CLT-010, CLT-010-1
Quantity73 units
DistributionFL ,GA, NJ, TN
Initiated2024-09-16
Classified2024-10-25
Recall numberZ-0231-2025
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)