RegWardenFDA facility compliance records
Recall

Class I — Cardinal Health 200, LLC

Recalling firmCardinal Health 200, LLC
Registered facilityView this company's 35 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass I
StatusOngoing
ReasonCertain lots of listed lots of Cardinal Health Monoject U-100 1 mL Insulin Syringe Luer-Lock with Tip Cap Soft Pack have demonstrated incompatibility with needleless Intravenous (IV) connectors. Cardinal Health recommends that they not be used for administration of IV push insulin via needleless connector.
ProductMonoject U-100 1 mL Insulin Syringe Luer-Lock with Tip Cap Soft Pack, REF 1188100777
Quantity371,326 units
DistributionNationwide
Initiated2024-09-25
Classified2024-11-08
Recall numberZ-0310-2025
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class I recall — Cardinal Health 200, LLC (Z-0310-2025).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/95487/

Not open yet

Alerts for Cardinal Health 200, LLC

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.