RegWardenFDA facility compliance records
Recall

Class II — Hobbs Medical, Inc.

Recalling firmHobbs Medical Inc
Registered facilityHOBBS MEDICAL, INC. (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonInjection solutions would not flow as intended through the tip of 25ga-sized needles. Investigations that some units contained debris from the manufacturing process within the inner sheath that blocked the ID of the needle.
ProductPosi-Stop Injection Needle, 23 Gauge, Catalog Number: 4706
Quantity6 units
DistributionWorldwide - US Nationwide distribution in the states of CT, GA, IN, MI, MN, MS, NH, NJ, NY, OH, TN, VT, WA, WI, WV and the country of Canada.
Initiated2024-10-04
Classified2024-11-15
Recall numberZ-0470-2025
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Hobbs Medical, Inc. (Z-0470-2025).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/95577/

Not open yet

Alerts for Hobbs Medical Inc

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.