RegWardenFDA facility compliance records
Recall

Class II — Micro-X Ltd.

Recalling firmMICRO-X LTD
Registered facilityMICRO-X LTD (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonDuring internal testing it was found that if the system is Ready to expose and the hand switch is then dropped into the stowage unit such that it lands on the prep/expose switch, the prep switch can be activated for a very short time such that the orange ring of death is displayed in the tube cooling area of the screen. The system then becomes unresponsive to kV/mAs changes, prep/expose presses. The only way to proceed is to suspend the patient, Exit Nexus DR and relogin and continue the exam.
ProductMICRO-X Rover Mobile X-ray System, # MXU-RV19
Quantity23
DistributionUS Nationwide Distribution: TX, CA, OH, ID, WA, AZ
Initiated2024-10-23
Classified2024-12-12
Recall numberZ-0666-2025
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Micro-X Ltd. (Z-0666-2025).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/95593/

Not open yet

Alerts for MICRO-X LTD

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.