RegWardenFDA facility compliance records
Recall

Class II — Zyno Medical LLC

Recalling firmZyno Medical LLC.
Registered facilityZyno Medical LLC. (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonPotential for the battery to deplete over time from self-discharge and internal parasitic leakage current despite being continuously plugged into AC mains.
ProductZ-800 Infusion System. Model/Catalog Number: Z-800, Z-800F, Z-800W, Z-800WF. The Z-800 Infusion system is intended to provide intravenous infusion of parenteral fluids, blood, and blood products to a patient under the direction or supervision of a physician or other certified healthcare professional.
Quantity34,994 units
DistributionThis product is distributed only in the United States.
Initiated2024-10-18
Classified2024-11-21
Recall numberZ-0506-2025
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Zyno Medical LLC (Z-0506-2025).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/95601/

Not open yet

Alerts for Zyno Medical LLC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.