Class II — Micro-X Ltd.

Recalling firmMICRO-X LTD
Registered facilityMICRO-X LTD (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonMobile x-ray system receiving/storage(operation) requirements to be followed: Temp:14 to 131(50 to 86) degrees F, RH:10% to 86%(30% to 60%) allow condensation dry time, Atmospheric Pressure:50kPa to 106 kPa(70kPa to 106kPa), Altitude:-102ft to 12,000 ft, (Maximum Gradient:9 degrees Fahrenheit), which mitigate risk of capacitor malfunction, unpredictable behavior, overheating, unintended start-up.
ProductMICRO-X Rover Mobile X-ray System, # MXU-RV35
Quantity6
DistributionUS: TX, PR, MD, CA
Initiated2024-10-22
Classified2024-11-27
Recall numberZ-0580-2025
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)