RegWardenFDA facility compliance records
Recall

Class II — Moximed, Inc.

Recalling firmMoximed, Inc.
Registered facilityMOXIMED, INC. (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonKnee implant may fracture due to a supplier-related issue with one component that has nonconforming geometry at the distal end of the absorber component that may result in a reduced wall thickness and microcracks, which may lead to new or worsening symptoms of discomfort, swelling / edema, pain or stiffness, which may necessitate removal of the device.
ProductMISHA Knee System Implant Small, Left, REF: 2-1001. MISHA Knee System Implant Small, Right, REF: 2-1002.
Quantity115
DistributionUS Nationwide distribution in the states of CA, OR, NY, OH, TX, CO, NY, VA.
Initiated2024-12-02
Classified2024-12-23
Recall numberZ-0762-2025
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Moximed, Inc. (Z-0762-2025).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/95887/

Not open yet

Alerts for Moximed, Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.