Recall

Class II — Spacelabs Healthcare, Inc.

Recalling firmSPACELABS HEALTHCARE, INC.
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonTelemetry receiver systems may experience a Windows proxy Issue causing the system to shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring. If a clinical event occurred during this period, staff may not detect until the system restarts; necessary clinical intervention could be delayed.
ProductXhibit Telemetry Receiver, Model 96280, with Service Manual. Intended to provide monitoring system with patient data.
Quantity2265
DistributionUS: OK, MN, IA, SC, NE, MI, NY, KY, GA, MO, MT, LA, WA, SD, CO, PA, WY, TX, ID, NJ, FL, OH, MS, OR, AK, IL, AR, KS, AL, TN, WI, VA, NM, CA, IN, RI, NV, MA, NC, ME, CT, AZ, MD, PR. OUS: Kuwait, Australia, Canada, Roumanie, Greece, Philippines, Spain, Italy, Hong Kong, Switzerland, Panama, Taiwan, Netherlands, Malaysia, Poland, Mexico, United Kingdom, Saudi Arabia, France, Czech Republic
Initiated2024-11-25
Classified2025-01-07
Recall numberZ-0809-2025
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Spacelabs Healthcare, Inc. (Z-0809-2025).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/95927/

Not open yet

Alerts for SPACELABS HEALTHCARE, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.