Class II — Murata Vios, Inc.
| Recalling firm | Murata Vios, Inc. |
|---|---|
| Registered facility | View this company's 2 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | A premature failure condition was identified with the L2050R pulse oximeter SpO2 sensor that connects to the L2050G Vios chest sensor device. When a failure occurs, a Pulse Ox Sensor Off warning condition will be incorrectly displayed by the system. |
| Product | Vios Monitoring System Bedside Monitor, Model BSM2050, Software Version Numbers 1.54.20 and 1.56.00 |
| Quantity | 44 systems |
| Distribution | US Nationwide distribution in the state of NY. |
| Initiated | 2023-06-06 |
| Classified | 2025-02-12 |
| Recall number | Z-1169-2025 |
| Initiation | Voluntary: Firm initiated |