RegWardenFDA facility compliance records
Recall

Class II — Augustine Temperature Management, LLC

Recalling firmAugustine Temperature Management, LLC
Registered facilityAugustine Temperature Management, LLC (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonThere is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, the connector cable on the mattress can experience the same type of deflection that caused the intermittent connection with the cables connecting to the ESU.
ProductHot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 89 cm (35"), REF US530, to prevent or treat hypothermia
Quantity3 units
DistributionUS Nationwide distribution in the states of TX, AL, NV; AU, NZ.
Initiated2024-12-16
Classified2025-02-20
Recall numberZ-1199-2025
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Augustine Temperature Management, LLC (Z-1199-2025).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/96155/

Not open yet

Alerts for Augustine Temperature Management, LLC

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.