RegWardenFDA facility compliance records
Recall

Class II — Kung Shin Plastics Co. Ltd.

Recalling firmKung Shin Plastics Co., Ltd.
Registered facilityKUNG SHIN PLASTICS CO., LTD. (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonThe machine side connector of the filter HME may be occluded by plastic which prevents proper airflow. The issue may lead to ventilation bag malfunction in which the user is unable to inflate/deflate properly during manual ventilation. Potential risks to health include oxygen deprivation, leading to respiratory distress and medical complications.
ProductCircuitGuard ThermoFlo Filter -Combined Heat and Moisture Exchanger, Bacterial Viral Filter w O2 Port and Elbow (Model/Catalogue number: SK200CPE / 6126)
Quantity50000 ea/ 500 cases
DistributionUS Nationwide distribution in the state of IL.
Initiated2025-01-09
Classified2025-03-04
Recall numberZ-1268-2025
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Kung Shin Plastics Co. Ltd. (Z-1268-2025).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/96224/

Not open yet

Alerts for Kung Shin Plastics Co., Ltd.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.