Class II — Maquet Cardiovascular, LLC
| Recalling firm | Maquet Cardiovascular LLC |
|---|---|
| Registered facility | View this company's 2 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | There have been at least 12 complaints from users as of March 3, 2025, reporting that they identified the ceramic C-Ring to be broken in half during use of the device. The ceramic C-Ring has a potential to break while in use. Maquet/Getinge determined that the issue was limited to the units of VH-4000 and VH-4001 manufactured with ceramic C-Ring, and the ceramic C-Ring was implemented in the Hemopro 2 devices from November 5, 2024, through February 14, 2025. |
| Product | The VASOVIEW HEMOPRO 2 Endoscopic Vessel Harvesting System is designed for use in conjunction with the 7 mm extended length endoscope. The Harvesting Cannula has four lumens which house the endoscope, C-Ring, distal lens washer tube, and Harvesting Tool for cutting and cauterizing vessel branches. The Harvesting Tool has two curved jaws. One of the jaws contains the heating elements for branch cutting and cauterization and spot cautery. An area near the tip of the convex side of the jaw is used for spot cautery. The activation toggle is used to control the jaws to activate the heating element… |
| Quantity | 22,244 units |
| Distribution | Worldwide distribution - US Nationwide and the countries of Mexico, United Arab Emirates, Singapore, India, Australia, Canada, Brazil, Hong Kong, Japan, South Korea, New Zealand. |
| Initiated | 2025-03-20 |
| Classified | 2025-04-18 |
| Recall number | Z-1622-2025 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Maquet Cardiovascular, LLC (Z-1622-2025).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/96439/
Alerts for Maquet Cardiovascular LLC
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?