Class II — Baxter Healthcare Corporation
| Recalling firm | BAXTER HEALTHCARE CORPORATION |
|---|---|
| Registered facility | View this company's 32 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | There were customer reports of devices which experienced battery-related fires. The investigation found that the devices in question were utilizing third-party, after-market batteries that are not approved by Baxter. |
| Product | Welch Allyn CONNEX Accessory Power Management Stand: |
| Quantity | 114,202 units |
| Distribution | Worldwide - US Nationwide distribution. |
| Initiated | 2025-03-24 |
| Classified | 2025-04-14 |
| Recall number | Z-1559-2025 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Baxter Healthcare Corporation (Z-1559-2025).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/96478/
Alerts for BAXTER HEALTHCARE CORPORATION
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.