Recall

Class II — Dexcom, Inc.

Recalling firmDEXCOM, INC.
Registered facilityView this company's 5 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonA bug in the 2.7.0 version of the G7 iOS App can cause the app to enter a state where it stops updating the estimated glucose value (EGV) and continues to show this stale EGV. If the user is unaware that the EGVs are not being updated and that glucose alerts will not trigger, there is potential for the missed detection of a hyperglycemic or hypoglycemic event or a treatment decision made based off incorrect data. There is no impact to transmitter communication with concurrently connected displays. A concurrently connected receiver, Apple Watch, and/or insulin pump will continue to receive EGVs directly from the transmitter when the app is in the problematic state. There is no impact to Automated Insulin Dosing (AID) therapy as the AID system continues to receive EGVs directly from the transmitter.
ProductDexcom G7 iOS Application, Ref: SW12300 used in conjunction with the Dexcom G7 Continuous Glucose Monitoring (CGM) system
Quantity61,355 users
DistributionWorldwide Distribution: US (nationwide) and OUS (foreign) to countries of: United Arab Emirates (AE), Austria (AT), Australia (AU), Belgium (BE), Bahrain (BH), Canada (CA), Switzerland (CH), Czech Republic (CZ), Germany (DE), Denmark (DK), Spain (ES), Finland (FI), France (FR), United Kingdom (GB), Hong Kong (HK), Ireland (IE), Israel (IL), Iceland (IS), Italy (IT), Jordan (JO), Korea, Repulic (KR…
Initiated2025-05-05
Classified2025-06-04
Recall numberZ-1906-2025
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Dexcom, Inc. (Z-1906-2025).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/96684/

Not open yet

Alerts for DEXCOM, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.