Class I — Ethicon Endo-Surgery Inc

Recalling firmETHICON ENDO-SURGERY, INC.
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass I
StatusOngoing
ReasonFirm has received an increase in reports regarding inadvertent instrument lockout during surgical procedures involving the device. If lockout occurs, the device will momentarily activate but will not cut or staple tissue, and additional steps will be required to open it and remove it from tissue. If user is unable to remove the device from the tissue, potential harms to patient include surgical delay, bleeding/hemorrhage, life threatening hemorrhage/hemorrhagic shock, and conversion to open therapy. Two adverse events received, including one patient death.
ProductEndopath Echelon Vascular White Reload for Advanced Placement Tip (35mm, 4 row); Product code VASECR35
Quantity678,526
DistributionUS distribution nationwide. OUS distribution to EMEA, LATAM, APAC, and Canada.
Initiated2025-04-22
Classified2025-05-20
Recall numberZ-1746-2025
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class I recall — Ethicon Endo-Surgery Inc (Z-1746-2025).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/96750/