Recall

Class II — Beckman Coulter, Inc.

Recalling firmBECKMAN COULTER, INC.
Registered facilityView this company's 22 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonInadequate preventative maintenance (PM) performed on the scales of the RPL5 fill line, led to splashing (cross-contamination) between the wells of testosterone reagent packs lot 538081, which could potentially result in up to a 34% increase or a 23% decrease in patient results.
ProductAccess Testosterone assay, Catalog Number 33560
Quantity70 units
DistributionUS State: FL. UAE, Vietnam
Initiated2025-05-05
Classified2025-06-02
Recall numberZ-1901-2025
InitiationVoluntary: Firm initiated