RegWardenFDA facility compliance records
Recall

Class II — Siemens Healthcare Diagnostics, Inc.

Recalling firmSIEMENS HEALTHCARE DIAGNOSTICS INC.
Registered facilityView this company's 12 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonContamination of the diluent by sodium hypochlorite (NaOCl) which may impact calibrator, quality control (QC), and patient results for all assays except for sodium, potassium, and chloride.
ProductProduct: Atellica CH Diluent - CONS 2 x 1.5L; Siemens Material Number (SMN): 11099300;
Quantity1,625 units (1450 US, 175 OUS)
DistributionWorldwide distribution - US Nationwide and the countries of Canada, Bahamas.
Initiated2025-05-22
Classified2025-06-25
Recall numberZ-2027-2025
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Siemens Healthcare Diagnostics, Inc. (Z-2027-2025).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/96952/

Not open yet

Alerts for SIEMENS HEALTHCARE DIAGNOSTICS INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.