RegWardenFDA facility compliance records
Recall

Class I — Tandem Diabetes Care, Inc.

Recalling firmTandem Diabetes Care, Inc.
Registered facilityView this company's 6 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass I
StatusOngoing
ReasonInsulin pumps manufactured with speakers from Revision A and B may malfunction ("Malfunction 16") which prevents audible alerts and causes the pumps to become inoperable and stop infusing insulin which could result in a hyperglycemic event, which may lead to injury and/or hospitalization.
ProductBattery Operated infusion pumps. t:slim X2 G5, Replacement, Refurbished, REF: 1000911; t:slim X2 GS Classic, Replacement, Refurbished, REF: 1003808; t: slim X2, Basal-IQ, mg/dl, REF: 1004219; Pump, t:slim X2, Clinical Use Only, REF: 1004484; Pump, t:slim X2, Control-IQ, REF: 1005611; Pump, t:slim X2, Control-IQ, mmoUL, Refurbished Replacement, REF: 1005613; Pump, t:slim X2, Control-IQ, mg/dl, REF: 1005615; Pump, t:slim X2, Control-IQ, mg/dl, Refurbished Replacement, REF: 1005617; Pump, t:slim X2, Basal-IQ 6.4, Refurbished Replacement, REF: 1006379; Pump, t:slim X2, Control-IQ 7.4, R…
Quantity171,750
DistributionWorldwide distribution. US nationwide including Puerto Rico, Canada, Czech Republic, Slovakia, Spain, United Kingdom, Belgium, Luxembourg, Australia, France, Italy, New Zealand, Finland, Sweden, Denmark, Norway, Saudi Arabia, Germany, Netherlands, Ireland, Portugal, South Africa, Switzerland
Initiated2025-07-21
Classified2025-08-27
Recall numberZ-2343-2025
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class I recall — Tandem Diabetes Care, Inc. (Z-2343-2025).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/97003/

Not open yet

Alerts for Tandem Diabetes Care, Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.