Recall

Class I — Hamilton Medical AG

Recalling firmHamilton Medical AG
Registered facilityView this company's 3 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass I
StatusOngoing
ReasonA mechanical defect on a printed circuit board (PCB) as a result of PCB separating process. This defect might cause the ventilator to either fail at startup (no patient involvement) or during the use of the device enter ambient state, resulting in an interruption of ventilation.
ProductHamilton C6 REF 160021 (ESM Board PN 10144626 UDI-DI code: 76300028PN160021ZW Hamilton C6 is a ventilator intended to provide positive pressure ventilatory support to adults and pediatrics and optionally infants and neonates.
Quantity1040 systems
DistributionUS Nationwide distribution.
Initiated2025-06-03
Classified2025-08-19
Recall numberZ-2328-2025
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class I recall — Hamilton Medical AG (Z-2328-2025).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/97079/

Not open yet

Alerts for Hamilton Medical AG

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.