RegWardenFDA facility compliance records
Recall

Class II — Boston Scientific Neuromodulation Corporation

Recalling firmBoston Scientific Corporation
Registered facilityView this company's 37 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonWire break(s) have occurred in rechargeable deep brain stimulation implantable pulse generators, implanted in the submuscular pectoral location. Wire breaks prevent successful stimulation, thus requiring device removal/replacement and users may experience high monopolar impedances, undesired sensation, loss of therapy, return of symptoms and/or Bluetooth connectivity challenges.
ProductVercise Deep Brain Stimulation Systems Surgical Implant Manual, Document Number: 92328632-09
Quantity25,260
DistributionWorldwide - US Nationwide distribution including in the states of MN, FL, CO, MD, NY, NE, PA, WI, AZ, TN, MO, TX, MI, MT, MA, WV, AL, VA, NC, CA, OR, WA, LA, OH, NJ, NH, SD, GA, CT, NV, UT, IN, ID, IL, ME, SC, DC, OK, KY, AK, HI, RI, AR, KS, NM and the countries of Argentina, Brazil, Chile, Ecuador, Colombia , Mexico, Canada, Japan, Algeria, Austria, Azerbaijan, Belgium, Bulgaria, Croatia, Denmark…
Initiated2025-07-08
Classified2025-09-12
Recall numberZ-2558-2025
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Boston Scientific Neuromodulation Corporation (Z-2558-2025).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/97144/

Not open yet

Alerts for Boston Scientific Corporation

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.