Class II — Boston Scientific Neuromodulation Corporation
| Recalling firm | Boston Scientific Corporation |
|---|---|
| Registered facility | View this company's 37 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | Wire break(s) have occurred in rechargeable deep brain stimulation implantable pulse generators, implanted in the submuscular pectoral location. Wire breaks prevent successful stimulation, thus requiring device removal/replacement and users may experience high monopolar impedances, undesired sensation, loss of therapy, return of symptoms and/or Bluetooth connectivity challenges. |
| Product | Vercise Deep Brain Stimulation Systems Surgical Implant Manual, Document Number: 92328632-09 |
| Quantity | 25,260 |
| Distribution | Worldwide - US Nationwide distribution including in the states of MN, FL, CO, MD, NY, NE, PA, WI, AZ, TN, MO, TX, MI, MT, MA, WV, AL, VA, NC, CA, OR, WA, LA, OH, NJ, NH, SD, GA, CT, NV, UT, IN, ID, IL, ME, SC, DC, OK, KY, AK, HI, RI, AR, KS, NM and the countries of Argentina, Brazil, Chile, Ecuador, Colombia , Mexico, Canada, Japan, Algeria, Austria, Azerbaijan, Belgium, Bulgaria, Croatia, Denmark… |
| Initiated | 2025-07-08 |
| Classified | 2025-09-12 |
| Recall number | Z-2558-2025 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Boston Scientific Neuromodulation Corporation (Z-2558-2025).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/97144/
Alerts for Boston Scientific Corporation
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.