RegWardenFDA facility compliance records
Recall

Class II — MEDLINE INDUSTRIES, LP - Northfield

Recalling firmMedline Industries, LP
Registered facilityView this company's 31 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonAffected kits contain recalled Stryker Color Cuff Sterile Disposable Tourniquet Cuffs. Stryker Instruments initiated this recall due to a potential manufacturing defect in which the flange may detach from the bladder at the welded connection point. This detachment compromises the cuff's ability to maintain adequate air pressure, potentially resulting in tourniquet failure and increased risk of patient hemorrhaging.
ProductThe Medline General Surgery Tray is customized to meet requirements of the hospitals and health care professionals utilizing them. The kit is provided sterile, single-use and includes a range of components dependent upon the specific procedure. These kits will be assembled using Medline and other manufacturers medical device components and drug components in some cases. Customers specify the kit contents, quantity, and placement of individual items in the kit.
Quantity1145
DistributionUS Nationwide distribution tin the states of CA, FL, IA, and TX.
Initiated2025-07-08
Classified2025-08-27
Recall numberZ-2452-2025
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — MEDLINE INDUSTRIES, LP - Northfield (Z-2452-2025).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/97371/

Not open yet

Alerts for Medline Industries, LP

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.