Class II — Cepheid
| Recalling firm | CEPHEID |
|---|---|
| Registered facility | View this company's 3 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | As a result of Post-Market half-time stability testing, false negative test result was observed for a positive sample type. |
| Product | Xpert SA Nasal Complete Catalog Number: GXSACOMP-10 UDI-DI code: 07332940000158 Xpert SA Nasal Complete kit; 10 test kit: A collection of reagents and other associated materials intended to be used for the direct qualitative and/or quantitative detection of nucleic acid from methicillin-resistant and methicillin-susceptible Staphylococcus aureus (MRSA/MSSA) bacteria present in a clinical specimen, using a nucleic acid technique |
| Quantity | 4,132 kits |
| Distribution | U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NH, NJ, NV, NY, OH, OK, OR, PA, PR, SC, TN, TX, UT, VA, WA, WI, WV, and WY. |
| Initiated | 2025-08-06 |
| Classified | 2025-10-30 |
| Recall number | Z-0414-2026 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Cepheid (Z-0414-2026).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/97497/
Alerts for CEPHEID
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.