Class II — ImaCor Inc.

Recalling firmImaCor Inc
Registered facilityIMACOR, Inc (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonProduction assembly error resulting in the potential for reversed articulation, where moving the lever forward caused the probe tip to anteflex rather than retroflex, and vice versa.
ProductClariTEE Miniaturized TEE probe. Model Number: CLT-010. Ultrasound miniaturized TEE probe for use with Mobile Ultrasound, ImaCor model ZHH.
Quantity45 units
DistributionUS Nationwide distribution in the states of South Carolina, New Jersey, Pennsylvania, Florida, and Georgia.
Initiated2025-08-29
Classified2025-10-03
Recall numberZ-0038-2026
InitiationVoluntary: Firm initiated