Recall

Class I — NOXBOX LTD

Recalling firmNOxBOX Limited
Registered facilityNOXBOX LIMITED (the firm's only registered facility)
ClassificationClass I
StatusOngoing
ReasonFluctuations may be more likely to occur if the total flow through the device sensor (used by the device to determine flow) is less than the minimum device requirement of 0.5 LPM. In addition, dose fluctuations have been observed when Bunnel LifePulse HFJV system or conventional ventilators have the following device settings: low service pressures generally less than 3.0 PSI, a pressure difference greater than 5.0cm H2O; or at breath per minute rates of 300 (5 Hz).
ProductNOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I
Quantity1667 units
DistributionDomestic: TN
Initiated2025-09-09
Classified2025-10-06
Recall numberZ-0028-2026
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class I recall — NOXBOX LTD (Z-0028-2026).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/97517/

Not open yet

Alerts for NOxBOX Limited

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.