RegWardenFDA facility compliance records
Recall

Class II — Beckman Coulter, Inc.

Recalling firmBECKMAN COULTER, INC.
Registered facilityView this company's 22 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonAutomated pipetting, diluting and specimen processing workstations for flow cytometric analysis device instructions for use list incorrect dead volumes: Cassette Type/List volume/correct volume, B/750/1582 microliters, and E/1300/2093 microliters. Device software does not rock Cassette E, used with 15mmX92mm Sarstedt tubes. May lead to incorrect diagnosis or error in patient management/treatment.
ProductCellMek SPS Sample Preparation System, REF: C44603. Software versions 2.2.122.0 and prior, and Instructions for Use (IFU) revisions AD and prior. These CellMek devices include Cassette E.
Quantity108 units
DistributionWorldwide - US Nationwide distribution in the states of AL, AR, FL, IL, IN, MA, ME, MI, MN, MO, NY, OR, TN, UT, WA and the countries Australia, Austria, Belgium, Brazil, Canada, China, France, Germany, Greece, Hong Kong, Israel, Italy, Japan, Korea, Republic of, Mexico, Morocco, Netherlands, Poland, Portugal, Puerto Rico, South Africa, Spain, Switzerland, United Kingdom of Great Britain and Northe…
Initiated2025-09-08
Classified2025-10-30
Recall numberZ-0412-2026
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Beckman Coulter, Inc. (Z-0412-2026).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/97531/

Not open yet

Alerts for BECKMAN COULTER, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.