RegWardenFDA facility compliance records
Recall

Class II — Diagnostica Stago, Inc.

Recalling firmDiagnostica Stago, Inc.
Registered facilityDIAGNOSTICA STAGO, INC. (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonA potential risk of underestimation of D-Dimer (D-Di) levels with the two listed lots.
ProductProduct: STA Liatest D-Di; REF: 00515;
Quantity12740 units
DistributionWorldwide distribution- US Nationwide and the country of Canada.
Initiated2025-09-12
Classified2025-10-16
Recall numberZ-0205-2026
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Diagnostica Stago, Inc. (Z-0205-2026).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/97596/

Not open yet

Alerts for Diagnostica Stago, Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.