Class II — Ferno-Washington Inc
| Recalling firm | Ferno-Washington, Inc. |
|---|---|
| Registered facility | FERNO-WASHINGTON, INC. (the firm's only registered facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | The fastening post hardware could loosen allowing the fastening post to detach from the cot |
| Product | FERNO POWER X2 POWERED AMBULANCE COT, Model/Catalog Number: 0015816; FOR USE IN GROUND TRANSPORTATION AMBULANCES. |
| Quantity | 138 units |
| Distribution | US Nationwide distribution in the states of AL, AR, AZ, CA, CO, FL, IS, IL, KY, LA, MN, MO, NC, ND, NE, NM, NY, OH, OK, PA, TN, TX, UT, VA, WI, WV. |
| Initiated | 2025-09-11 |
| Classified | 2025-10-10 |
| Recall number | Z-0133-2026 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Ferno-Washington Inc (Z-0133-2026).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/97606/