Class I — Cook Medical Incorporated
| Recalling firm | COOK MEDICAL |
|---|---|
| Registered facility | COOK MEDICAL LLC (the firm's only registered facility) |
| Classification | Class I |
| Status | Ongoing |
| Reason | Affected devices may contain PTFE coating scrapings. Scrapings could be released during device deployment, potentially causing intravascular embolization. |
| Product | Zenith Alpha 2 Thoracic Endovascular Graft, Catalog Prefix ZTA2- Reference Part Numbers ZTA2-P-40-117-W ZTA2-P-40-167 ZTA2-P-40-167-W ZTA2-P-40-217-W ZTA2-P-42-121 ZTA2-P-42-121-W ZTA2-P-42-173 ZTA2-P-42-173-W ZTA2-P-42-225 ZTA2-P-42-225-W ZTA2-P-44-125 ZTA2-P-44-125-W ZTA2-P-44-179 ZTA2-P-44-179-W ZTA2-P-44-233 ZTA2-P-44-233-W ZTA2-P-46-125 ZTA2-P-46-125-W ZTA2-P-46-179 ZTA2-P-46-179-W ZTA2-P-46-233-W ZTA2-PT-40-36-167 ZTA2-PT-40-36-167-W ZTA2-PT-40-36-217 ZTA2-PT-40-36-217-W ZTA2-PT-42-38-173 ZTA2-PT-42-38-173-W ZTA2-PT-42-38-225 ZTA2-PT-42-38-225-W ZTA2-PT-4… |
| Quantity | 695 |
| Distribution | Domestic distribution to AL AR AZ CA CT DC DE FL GA IL IN LA MA MD MI MN MO NC NJ NY OH OR PA SC TN TX UT VA WA WI WV International distribution to Switzerland, Germany, Italy, Poland, and Sweden |
| Initiated | 2025-09-18 |
| Classified | 2025-10-10 |
| Recall number | Z-0043-2026 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class I recall — Cook Medical Incorporated (Z-0043-2026).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/97615/
Alerts for COOK MEDICAL
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?