Class I — BALT USA, LLC

Recalling firmBalt USA, LLC
Registered facilityBALT USA, LLC (the firm's only registered facility)
ClassificationClass I
StatusOngoing
ReasonDue to an identified manufacturing defect involving coating material lifting within the catheter's distal inner diameter/lumen.
ProductBrand Name: MEGA Ballast Distal Access Platform Product Name: Mega Ballast Model/Catalog Number: MEGABT09180, MEGABT09190, MEGABT091100, MEGABT091110 Software Version: N/A Product Description: The Mega Ballast 091 is designed with a semi-rigid proximal shaft and a highly flexible distal shaft. The design facilitates the advancement of the Mega Ballast to access the tortuous vasculature over a guidewire and a respective dilator. The outer surface of the Mega Ballast is coated with a hydrophilic coating to facilitate lubricity from tip to 40 cm proximal. The proximal end of the Mega Ballast …
Quantity207 catheters
DistributionUS distribution to states of: AL, CA, CO, CT, FL, IL, KY, MA, NY, TN, TX, and UT.
Initiated2025-09-25
Classified2025-10-24
Recall numberZ-0188-2026
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class I recall — BALT USA, LLC (Z-0188-2026).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/97661/

Not open yet

Alerts for Balt USA, LLC

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.