Class II — Medistim Asa
| Recalling firm | Medistim ASA |
|---|---|
| Registered facility | View this company's 2 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | Printer Support Module (PSM) may be subject to insufficient soldering, which may cause loosening of the wires and system to shut down. Affected devices need to have their Printer Support Modules replaced. |
| Product | Brand Name: Medistim MiraQ System Product Name: Medistim MiraQ System Model/Catalog Number: MQC04001, MQC14001, MQU14011, MQV12011 Software Version: Not applicable Product Description: Ultrasound device for intraoperative use Component: No |
| Quantity | 14 |
| Distribution | Domestic distribution to CO, MI, MN, TN, TX. International distribution to Australia, Austria, Germany, Finland, India, Japan, Russia, China, Saudi Arabia. |
| Initiated | 2025-09-26 |
| Classified | 2025-10-29 |
| Recall number | Z-0410-2026 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Medistim Asa (Z-0410-2026).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/97707/
Alerts for Medistim ASA
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?