| Reason | An internal investigation has determined that due to defective sub-components within the gantry assembly, a z-axis CAD movement motion error may occur and cause an analyzer to enter a Not Ready or Paused state, leading to a delay in obtaining results. |
| Product | Beckman Coulter UniCel DxI 600, Part Numbers: A71460, A30260; Photometric Chemistry Analyzer |
| Quantity | 80 units |
| Distribution | Worldwide |
| Initiated | 2025-10-01 |
| Classified | 2025-11-04 |
| Recall number | Z-0428-2026 |
| Initiation | Voluntary: Firm initiated |