RegWardenFDA facility compliance records
Recall

Class II — Pivotal Health Solutions, Inc.

Recalling firmPivotal Health Solutions, Inc.
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonThe heat function may become damaged over time due to patients pressing on the areas with heat function components. Damage to components can lead to overheating which can lead to smoldering and melting of the vinyl cushion.
ProductQuantum Intersegmental Table, Model/Catalog Number: AMQ400H; heated massage table
Quantity464 units
DistributionDistribution to US nationwide, Canada and Japan.
Initiated2025-10-09
Classified2025-11-24
Recall numberZ-0588-2026
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)