Class II — Inpeco S.A.
| Recalling firm | Inpeco SA |
|---|---|
| Registered facility | INPECO SA (the firm's only registered facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | The Aliquoter Module (mALX) may generate secondary sample tubes without labels or with mismatched labels which may result in delayed results or erroneous results. This error occurs if the operators open the printer head of mALX to perform error recoveries or maintenance procedures and do not empty the printer tubes buffer to remove all the secondary sample tubes. |
| Product | FlexLab (FLX); Version: FLX-217-10; |
| Quantity | 1 unit (OUS only) |
| Distribution | US Domestic: MI. OUS International: Australia, Finland, France, Germany, Italy, Portugal, Spain, and UK. |
| Initiated | 2025-10-30 |
| Classified | 2025-11-25 |
| Recall number | Z-0595-2026 |
| Initiation | Voluntary: Firm initiated |