Class II — AGFA Healthcare Corp.
| Recalling firm | AGFA HEALTHCARE N.V. |
|---|---|
| Registered facility | View this company's 4 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | It is possible to exceed the maximum allowed Air Kerma rate (AKR) of 88mGy/min for pulsed fluoro exams. This can happen in case the framerate is increased on the NX workstation when ABS is not enabled. |
| Product | Brand name: AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic Common Name: DR 800 |
| Quantity | 35 |
| Distribution | U.S. Nationwide distribution. |
| Initiated | 2025-11-18 |
| Classified | 2025-12-22 |
| Recall number | Z-0923-2026 |
| Initiation | FDA Mandated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — AGFA Healthcare Corp. (Z-0923-2026).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/98101/
Alerts for AGFA HEALTHCARE N.V.