Class II — Securitas Healthcare LLC
| Recalling firm | Securitas Healthcare LLC. |
|---|---|
| Registered facility | AeroScout, Ltd. (the firm's only registered facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | Securitas Healthcare has become aware of a firmware issue in the Call Stations in the Affected Lots that may cause Call Stations in the Affected Lots to (1) not transmit a low battery alert to the Arial Wireless Emergency Call System prior to battery failure, or (2) not transmit a low battery alert to the Arial Wireless Emergency Call system at least seven (7) days prior to the battery failing. |
| Product | Arial 900 MHz Call Station, Model Number CSK200-900MR, with Firmware version: 1.6.3 |
| Quantity | 1898 units |
| Distribution | Worldwide distribution -US Nationwide and the country of Canada. |
| Initiated | 2025-12-08 |
| Classified | 2026-01-08 |
| Recall number | Z-1032-2026 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Securitas Healthcare LLC (Z-1032-2026).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/98122/
Alerts for Securitas Healthcare LLC.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.