Class II — FUJIFILM Healthcare Americas Corporation

Recalling firmFUJIFILM Healthcare Americas Corporation
Registered facilityFUJIFILM Healthcare Americas Corporation (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonIt was found that the contacts of the electromagnetic contactors used inside the cabinet of the X-ray high-voltage generator were welded, and the current might continue to be supplied to the power supply circuit even if the power of the device is cut off. As a result, it cannot be denied that the circuit in question overheats and damages the surrounding parts. It may also cause smoke.
ProductFDR Visionary Suite; Model Number: CH-200; Version: (1) 566-16130-23, (2) 566-16130-31, (3)
Quantity16 units
DistributionUS Nationwide distribution in the states of AK, CA, CT, FL, IL, NH, OH, WI.
Initiated2026-01-09
Classified2026-02-05
Recall numberZ-1299-2026
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — FUJIFILM Healthcare Americas Corporation (Z-1299-2026).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/98124/

Not open yet

Alerts for FUJIFILM Healthcare Americas Corporation

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.