Class II — Boston Scientific Corporation
| Recalling firm | Boston Scientific Corporation |
|---|---|
| Registered facility | View this company's 37 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | Boston Scientific is initiating the removal of certain batches of Coyote" Over-the-Wire (OTW) Percutaneous Transluminal Angioplasty (PTA) Balloon Dilatation Catheter devices due to the potential for inflation/deflation difficulty during use in the peripheral vasculature. An internal investigation found that a manufacturing process variation resulted in catheters that may not properly inflate/deflate. This was limited to Coyote OTW batches manufactured from 12Sep2025 through 13Oct2025. The processing variation has since been addressed. No other material numbers or batches are impacted by this removal. |
| Product | Boston Scientific Coyote OVER-THE-WIRE PTA Balloon Dilation Catheter, 4F, 1) 2.5mm x120mm, 150 cm, UPM/Model No. REF H74939186251210; 2) 3.5mm x120mm, 150 cm, UPM/Model No. REF H74939186351210; Catheter, Angioplasty, Peripheral, Transluminal |
| Quantity | 21 units |
| Distribution | US Nationwide distribution in the states of MA, WV, OH, CA, MN, NY, VA, IL TX, FL, IL. |
| Initiated | 2025-12-03 |
| Classified | 2026-01-09 |
| Recall number | Z-1042-2026 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Boston Scientific Corporation (Z-1042-2026).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/98149/
Alerts for Boston Scientific Corporation
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.