Class II — ICU Medical, Inc.
| Recalling firm | ICU MEDICAL, INC. |
|---|---|
| Registered facility | View this company's 6 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | Needle free access device intended as accessory to vascular access used in hemodialysis or accessory to intravascular administration sets has a silicone seal which may bulge at the top, separate from the body, tear; may result in fluid leaks, occluded fluid flow, inability to inject/withdraw blood product, delayed/interrupted therapy, exposure to biological contaminants, air infused into the body |
| Product | Tego Connector: REF: 011-D1000, 01C-D1000, 055-D1000, D1000; Conector Tego: REF: LAT-D1000 |
| Quantity | 14,684,607 items |
| Distribution | Worldwide - US Nationwide distribution including in the states of FL, AZ, RI, TN, KS, NJ, NY, VA, SC, GA, CA, MS, TX, OH, NC, MA, WA, IL, PA, NV, NH, LA, MI, NE, ME, MN, NM, OK, MD, IA, OR, SD, WI, MO, AL, WY, IN, KY, DC, UT, AR, WV, AK, ND, CO and the countries of BE, DE, CO, ES, AU, AR, SA, SI, IT, LK, MY, AT, FI, GB, TW, MX, NL, IE, CA, GR, ZA, SE, SK, NZ, LT, CH, HR, FR, JO, PE, SG, BR, PL, RO… |
| Initiated | 2025-12-19 |
| Classified | 2026-01-16 |
| Recall number | Z-1111-2026 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — ICU Medical, Inc. (Z-1111-2026).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/98150/
Alerts for ICU MEDICAL, INC.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.