RegWardenFDA facility compliance records
Recall

Class II — Staar Surgical AG

Recalling firmSTAAR SURGICAL Company
Registered facilityView this company's 3 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonToric Implantable Collamer for correction/reduction of myopia in phakic eyes with astigmatism were labeled as 12.1 mm length but are 12.6 mm length lenses, so there is the potential for excessive vault, which may cause: Significant reduction of irido-corneal angles, Significant shallowing of anterior chamber, Patient discomfort, Elevated IOP/Angle closure, Unreactive pupil, Lens exchange/removal.
ProductEVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5
Quantity7
DistributionInternational distribution to the countries of India, Iran, Korea, Saudi Arabia.
Initiated2026-01-21
Classified2026-03-02
Recall numberZ-1480-2026
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Staar Surgical AG (Z-1480-2026).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/98397/

Not open yet

Alerts for STAAR SURGICAL Company

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.