Recall

Class II — Reflexion Medical, Inc.

Recalling firmRefleXion Medical, Inc.
Registered facilityREFLEXION MEDICAL, INC. (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonDue to a radiotherapy medical system software defect, when performing angular roll corrections followed by a repeat localization, the roll corrections applied in the initial localization scan are not carried through to treatment delivery leading to the treatment plan dose being delivered to the incorrect location, potentially a displacement of 5mm or higher, with a 10% to 20% underdose.
ProductRefleXion X1 Radiotherapy System Model RXM1000, Part #800-00006-00
Quantity12
DistributionUS: CA, TX, PA, CT, NJ, OR, LA, OH
Initiated2026-02-09
Classified2026-03-27
Recall numberZ-1656-2026
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Reflexion Medical, Inc. (Z-1656-2026).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/98518/

Not open yet

Alerts for RefleXion Medical, Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.