Class II — Beckman Coulter, Inc.
| Recalling firm | BECKMAN COULTER, INC. |
|---|---|
| Registered facility | View this company's 22 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | Beckman Coulter has identified that Access Total T4 Calibrator lots (PN 33805) exhibit a negative bias on DxI 600/800 instruments, with slopes outside product specifications. A false low patient result could cause a patient to undergo unnecessary additional diagnostic testing. A patient may be treated for a disorder not present or receive delayed treatment for hyperthyroidism. For a pregnant patient being monitored for thyroid hormone supplementation, incorrect treatment or dose may be used. A delayed patient results could cause repeat blood draw with associated complications. |
| Product | Access Total T4 Calibrator, Catalog No. 33805 |
| Quantity | 2068 US, 2830 OUS |
| Distribution | Worldwide distribution - US Nationwide and the countries of Argentina, Austria, Bangladesh, Bolivia (Plurinational State of), Bosnia and Herzegovina, Brazil, Canada, China, Cte d'Ivoire, Croatia, Czechia, Denmark, Egypt, El Salvador, Ethiopia, Georgia, Germany, Greece, Hong Kong, India, Indonesia, Ireland, Italy, Japan, Kazakhstan, Korea, Republic of, Kosovo, Lebanon, Libya, Malawi, Mexico, Myanm… |
| Initiated | 2026-03-16 |
| Classified | 2026-04-22 |
| Recall number | Z-1922-2026 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Beckman Coulter, Inc. (Z-1922-2026).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/98591/