Class II — Dexcom, Inc.
| Recalling firm | DEXCOM, INC. |
|---|---|
| Registered facility | View this company's 5 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | A software defect in which a low-priority file I/O check blocks the higher-priority thread responsible for processing estimated glucose values (EGVs) and trend arrows. When the app is running in the background, EGV notifications may queue up and then "replay" sequentially when the user brings the app to the foreground, causing the Glucose Compass to flicker through previous EGVs. In more severe instances, the app may perceive signal loss and deploy signal loss mitigations. This delay in EGV and alert processing creates a risk that users may miss detection of a hypoglycemic or hyperglycemic event, or make treatment decisions based on outdated glucose information. |
| Product | Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM iOS App Model/Catalog Number: SW12300 Software Version: Versions 2.9.0 through 2.11.2 Product Description: Dexcom G7 CGM iOS App Component: Dexcom G7 Continuous Glucose Monitoring System, Dexcom G7 15 Day Continuous Glucose Monitoring System |
| Quantity | 58582 |
| Distribution | SW12300 & SW13355 Dexcom G7 iOS and watchOS app versions 2.9.0 through 2.11.2: Worldwide - US Nationwide including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Austria, Australia, Belgium, Bahrain, C… |
| Initiated | 2026-04-14 |
| Classified | 2026-05-11 |
| Recall number | Z-2163-2026 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Dexcom, Inc. (Z-2163-2026).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/98639/
Alerts for DEXCOM, INC.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.