Class II — The Binding Site Group, Ltd.
| Recalling firm | THE BINDING SITE, INC. |
|---|---|
| Registered facility | THE BINDING SITE, INC. (the firm's only registered facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | It was reported that plates 2 and 3 were marked as passed , although it appeared that QC had not been performed on these plates, despite samples being measured. The potential clinical hazards include false or missed detection of M-protein (false positive); and falsely elevated or decreased M-protein concentration. |
| Product | Brand Name: EXENT Analyser Product Name: EXENT Analyser Model/Catalog Number: IE800.A Software Version: 1.0.20 Product Description: The EXENT Analyser is an automated analyser intended for the qualitative and quantitative in vitro measurement of analytes in human body fluids used in conjunction with the EXENT assays. |
| Quantity | 1 |
| Distribution | US Nationwide distribution. |
| Initiated | 2026-04-16 |
| Classified | 2026-06-09 |
| Recall number | Z-2362-2026 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — The Binding Site Group, Ltd. (Z-2362-2026).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/98830/
Alerts for THE BINDING SITE, INC.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.