RegWardenFDA facility compliance records
Recall

Class I — Insulet Corporation

Recalling firmINSULET CORPORATION
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass I
StatusOngoing
ReasonExternal soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary failure mode is pump under-delivery due to loss of insulin to an external leak. If insulin is not delivered properly, users may experience high blood glucose levels due to under-delivery of insulin. In the most severe cases, prolonged and persistent high blood glucose levels can lead to diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or even death.
ProductOmnipod DASH Pod. Includes the below Model/REF Numbers: 1. POD-BLE-I1-525 (5-Pack), POD-BLE-I1-520 (Single Pod). 2. POD-BLE-I1-529 (10-Pack), POD-BLE-I1-520 (Single Pod). 3. BLE-I1-529 (10-Pack), BLE-I1-520 (Single Pod). 4. BLE-C1-529 (10-Pack), BLE-C1-520 (Single Pod).
Quantity1,790,784 units
DistributionWorldwide distribution - US Nationwide and the countries of Austria, Australia, Belgium, Canada, Switzerland, Germany, Denmark, Finland, France, UK, Greece, Israel, Italy, Kuwait, Leichtenstein, Netherlands, Norway, Saudi Arabia, Sweden and Turkey.
Initiated2026-05-20
Classified2026-06-30
Recall numberZ-2604-2026
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class I recall — Insulet Corporation (Z-2604-2026).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/99034/

Not open yet

Alerts for INSULET CORPORATION

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.