Class II — Accuray Incorporated

Recalling firmAccuray Incorporated
Registered facilityAccuray Incorporated (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonIn certain cases, a collimator may not fully dock, which triggers a system interlock and prevents motion from the operator console. If personnel then enter the treatment room and initiate motion using the in-room control pendant, the collimator may release and fall from the housing.
ProductCyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator Changer; Linear Accelerator
Quantity48 units
DistributionWorldwide distribution - US Nationwide and the countries of China, Colombia, Czech Republic, France, Hong Kong, Japan, Korea, Republic of, Latvia, Mexico, Myanmar, Saudi Arabia, Taiwan, Turkey, Ukraine.
Initiated2026-06-16
Classified2026-07-01
Recall numberZ-2640-2026
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Accuray Incorporated (Z-2640-2026).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/99075/

Not open yet

Alerts for Accuray Incorporated

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.