Class II — Tobii Dynavox Llc

Recalling firmTobii Dynavox LLC
Registered facilityDYNAVOX Group AB (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonUnder certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure.
ProductTD Navio Maxi Speech Generating Device Model/Catalog Number: 1000799
DistributionUS Nationwide distribution.
Initiated2026-07-08
Classified2026-08-14
Recall numberZ-2960-2026
InitiationVoluntary: Firm initiated